Expand Up @@ -392,6 +392,8 @@ the first few rings.
:align: center
:width: 80%

*Points generated for the innermost ring after selecting four points*

After clicking on "Select Points", click somewhere on the innermost
ring. This triggers the PyFAI Massif module, which automatically detects
other points on the Debye-Scherrer ring that are contiguous to the
Expand All @@ -411,9 +413,10 @@ generate, using the ring pull-down menu.
:align: center
:width: 80%

In this example, the pull-down menu was set to "Ring6". When enough
rings have been defined, click "Calibrate" again to produce a more
accurate refinement.
*Autogenerated rings after selecting "Ring6" on the pull-down menu*

When enough rings have been defined, click "Calibrate" again to produce
a more accurate refinement.

The "Plot Cake" button can be used to generate a "cake" plot, in which
all the powder rings, which are plotted against polar angle, should fall on vertical lines.
Expand All @@ -422,12 +425,15 @@ all the powder rings, which are plotted against polar angle, should fall on vert
:align: center
:width: 80%

*Cake Plot*

This can be used to determine whether the calibration is sufficiently
good over the entire angular range of the detector. If there is evidence
of distortions at higher polar angle, it may be necessary to
autogenerate more rings before an additional calibration.

When the calibration is satisfactory, click "Save" to save both the
powder calibration data and parameters to the configuration file. This
process should be repeated for each entry, after which the dialog can be
closed.
powder calibration data and parameters to the configuration file. The
calibration parameters can also be saved to a PONI file, using the
"Export Calibration" button. This process should be repeated for each
entry, after which the dialog can be closed.