Skip to content

Commit

Permalink
Merge pull request #21 from ivalaginja/cleanup
Browse files Browse the repository at this point in the history
More typos and clarifications
  • Loading branch information
ivalaginja authored Oct 10, 2024
2 parents 1f5c708 + 4a43e93 commit 4ec32a8
Show file tree
Hide file tree
Showing 5 changed files with 14 additions and 15 deletions.
6 changes: 3 additions & 3 deletions _pages/about.md
Original file line number Diff line number Diff line change
Expand Up @@ -36,7 +36,7 @@ easily to conferences and outreach events.

The BabyCAT project was designed in 2019 by a team of scientists at the
[Russell B. Makidon Optics Laboratory](https://www.stsci.edu/stsci-research/research-topics-and-programs/russell-b-makidon-optics-laboratory)
at the [Space Telescope Science Institite](https://www.stsci.edu/).
at the [Space Telescope Science Institute](https://www.stsci.edu/).
Its goal is to make advanced optical techniques more accessible to a wider audience.

The first ever BabyCAT coronagraph demonstration bench was built in 2019 and has since been used in various educational
Expand All @@ -49,8 +49,8 @@ who worked with the RMOL team and STScI's Office of Public Outreach (OPO) during
"High-contrast imager for Complex Aperture Telescopes" testbed. Since the demo coronagraph was supposed to illustrate a
mini version of HiCAT, the name "BabyCAT" seemed immediately evident.

The second demo bench, BabyCAT 2.0, was also assembled at the Makidon lab and later delivered to ExEP at JPL. Its
upgraded design is the one that is presented on this website.
The second demo bench, BabyCAT 2.0, was also assembled at the Makidon lab and later delivered to ExEP at NASA's Jet
Propulsion Laboratory (JPL). Its upgraded design is the one that is presented on this website.

## Where This Setup Is Used

Expand Down
18 changes: 9 additions & 9 deletions _pages/assembly.md
Original file line number Diff line number Diff line change
Expand Up @@ -40,7 +40,7 @@ In this setup, the three optical elements 1-3 (OE1, OE2, OE3) are spherical mirr

![Optical Layout](/assets/images/BabyCAT_unfolded.png){: .align-center}

Below, you can see close-up images of the reflective focal-plane mask (left) and the Lyot stop iris (right) on their respective magnetic
Below, you can see close-up images of the reflective focal-plane mask (FPM, left) and the Lyot stop (LS) iris (right) on their respective magnetic
mounts.

<img src="{{ site.url }}{{ site.baseurl }}/assets/images/fpm_ls.png" alt="FPM and LS" width="600"/>{: .align-center}
Expand All @@ -58,14 +58,14 @@ The below is a schematic of the general layout of the demo coronagraph.
[Download PDF](https://github.com/ivalaginja/babycat/tree/main/assets/pdfs/assembly_outline.pdf){: .btn .btn--danger}

The alignment of the demo coronagraph requires some familiarity with standard optical alignment techniques and the use
of standard optics mounts, as well as good judgment as to when PSFs and images are in good focus. For cost reasons, most
of standard optics mounts, as well as good judgment as to when point-spread functions (PSFs) and images are in good focus. For cost reasons, most
mounts in BabyCAT do not have adjustment actuators. As a result, adjustments have to be made by carefully sliding mounts
across the breadboard and by sliding posts up and down in post holders. Care is required when performing these movements
and when screwing down mounts in their final location.

The final goal is a demonstration that is visually instructive regarding the operation of a coronagraph for exo-planet
The final goal is a demonstration that is visually instructive regarding the operation of a coronagraph for exoplanet
imaging. Precision alignments may not strictly be necessary to achieve this. However, experience has shown that some
care and finesse is needed to align the system well. Otherwise, there will be appreciable aberrations (primarily
care and finesse are needed to align the system well. Otherwise, there will be appreciable aberrations (primarily
astigmatism) visible in the system PSF and the “star” light will not be effectively rejected. It is also possible to
have a system that is nearly aligned, but a mechanical interference presents itself when attempting to, for example,
perform final focusing on the Lyot stop. Multiple iterations of the alignment procedure may be required to achieve good
Expand Down Expand Up @@ -99,7 +99,7 @@ helps reduce angles of incidence and minimize astigmatism. (Note that in an alre
5. Adjust the lateral position of the FPM to align the beam with the pinhole.
6. Adjust pupil camera such that it is perpendicular to the beam path. Use a transmissive paper to
establish beam path and ease centering on the camera.
7. Adjust pupil camera to establish best focus, using the sharpness of PSF edges as a guide.
7. Adjust pupil camera to establish best focus, using the sharpness of pupil edges as a guide.

### Refocusing the focal-plane mask

Expand All @@ -108,15 +108,15 @@ disassembled and repositioned. Note that moving the FPM changes the conjugates (
spherical mirror and final imaging optic.

1. Disassemble the imaging backend.
2. Insert a pickoff (flat) mirror and re-imaging lens to acquire an image of the back of the FPM.
3. The camera can be brought to focus by finding the plane in which the PSF is best-focused (which will be conjugate to the best focus of the FPM.)
2. Insert a pick-off (flat) mirror and re-imaging lens to acquire an image of the back of the FPM.
3. The camera can be brought to focus by finding the plane in which the PSF is best-focused (which will be conjugate to the best focus of the FPM).
4. Adjust the FPM position to maintain centering and bring the edges into focus.
5. Re-align the imaging backend (may take ~ 1 hour).
5. Re-align the imaging backend (may take ~1 hour).

## Assembly images

Below you can find a couple of images of the assembled BabyCAT and BabyCAT2 benches. The assembly with a grey frame is
BabyCAT, and the assembly with a black frame is the evolved BabyCAT2, on which the parts list on this website is based on.
BabyCAT, and the assembly with a black frame is the evolved BabyCAT2, on which the parts list of this website is based on.

Keep in mind that the assembly images are for reference only and may not be the most up-to-date version of the benches.
For example, the images of BabyCAT show the green laser listed in the parts list, while the images of BabyCAT2 show a
Expand Down
3 changes: 1 addition & 2 deletions _pages/parts-list.md
Original file line number Diff line number Diff line change
Expand Up @@ -7,8 +7,7 @@ sidebar:
nav: "parts"
---

This page lists all the parts required to build the portable coronagraph. The list is divided into several categories
listed in the navigation bar on the left.
This page lists all the parts required to build the portable coronagraph.

To build a fully functional coronagraph bench, you will need all components from the two sections
[Optical Components](#optical-components) and [Screws and tools](#screws-and-tools). Note how for the screws, you could
Expand Down
Binary file modified assets/pdfs/assembly_outline.pdf
Binary file not shown.
2 changes: 1 addition & 1 deletion index.md
Original file line number Diff line number Diff line change
Expand Up @@ -8,7 +8,7 @@ header:
actions:
- label: "Get parts list"
url: /parts-list/
caption: "Photo credit: [**R. Soummer**]"
caption: "Photo credit: **R. Soummer**"
excerpt: "A portable coronagraph bench."
intro:
- excerpt: 'Welcome to the BabyCAT website! BabyCAT is a portable coronagraph bench designed to interactively demonstrate coronagraphy. You can download a full parts list and build your own BabyCAT - make sure to let us know if you do!'
Expand Down

0 comments on commit 4ec32a8

Please sign in to comment.