Other QC stats #1

Open
cookpa opened this issue May 20, 2024 · 2 comments
Open

Other QC stats #1

cookpa opened this issue May 20, 2024 · 2 comments

Comments

@cookpa
Copy link
Contributor

cookpa commented May 20, 2024

If there's interest, could look at other QC stats:

  • ref distortion (make sure the first two b=0 have no in-plane distortion), currently we only output (ref distortion) / (eddy distortion)

  • b=0 signal drift - might be tricky because the test protocol will be quite short, but could plot number anyway for tracking

  • Phantom ADC

  • Others?

@cookpa
Copy link
Contributor Author

cookpa commented May 20, 2024

The signal falls off pretty sharply with b-value so if we are interested in computing ADC, it might be wise to get some measurements at b=500, I'd probably replace the b=1000 shell with b=500

@cookpa
Copy link
Contributor Author

cookpa commented May 20, 2024

Actually, the b=1000 shell might be OK - it's faint but still distinguishable from background.